Autoria: |
AMARAL, C. D. B. do; MACHADO, R. C.; BARROS, J. A. V. de A.; VIRGILIO, A.; SCHIAVO, D.; NOGUEIRA, A. R. de A.; NOBREGA, J. de A. |
Afiliação: |
Clarice Dias Britto do Amaral, UFSCar; Raquel Cardoso Machado, UFSCar; Juan Antônio Vázquez de Almeida Barros, Agilent Technologies; Alex Virgilio, UFSCar; Daniela Schiavo, UFSCar; ANA RITA DE ARAUJO NOGUEIRA, CPPSE; Joaquim de Araújo Nóbrega, UFSCar. |
Título: |
Determination of rare earth elements in geological and agricultural samples by ICP-OES. |
Ano de publicação: |
2017 |
Fonte/Imprenta: |
Espectroscopy, v. 32, n. 10, p. 2-6, 2017. |
Idioma: |
Inglês |
Conteúdo: |
The selection of an analytical method featuring multielemental capability, wide linear range, and that stands for an elevated content of total dissolved solids (TDS) would be ideal for rare earth elements (REEs) determination in complex samples. This work targets the determination of Ce, Dy, Er, Eu, Gd, Ho, La, Lu, Nd, Pr, Sc, Sm, Tb, Th, Tm, Y, and Yb in geological samples, fertilizer, and agricultural gypsum samples using inductively coupled plasma?optical emission spectrometry (ICP-OES). Samples were prepared using microwave-assisted acid digestion and the accuracy was evaluated by addition and recovery experiments for spiked samples at three concentration levels. A careful choice of the plasma viewing position and appropriate dilution factors led to accurate and sensitive determinations of REEs. The proposed procedure using external calibration presented good accuracy and sensitivity. |
Palavras-Chave: |
Geological material; Spectrometry. |
Thesaurus Nal: |
rare earth elements. |
Categoria do assunto: |
X Pesquisa, Tecnologia e Engenharia |
Marc: |
LEADER 01568nam a2200217 a 4500 001 2079572 005 2017-11-13 008 2017 bl uuuu u00u1 u #d 100 1 $aAMARAL, C. D. B. do 245 $aDetermination of rare earth elements in geological and agricultural samples by ICP-OES.$h[electronic resource] 260 $aEspectroscopy, v. 32, n. 10, p. 2-6, 2017.$c2017 520 $aThe selection of an analytical method featuring multielemental capability, wide linear range, and that stands for an elevated content of total dissolved solids (TDS) would be ideal for rare earth elements (REEs) determination in complex samples. This work targets the determination of Ce, Dy, Er, Eu, Gd, Ho, La, Lu, Nd, Pr, Sc, Sm, Tb, Th, Tm, Y, and Yb in geological samples, fertilizer, and agricultural gypsum samples using inductively coupled plasma?optical emission spectrometry (ICP-OES). Samples were prepared using microwave-assisted acid digestion and the accuracy was evaluated by addition and recovery experiments for spiked samples at three concentration levels. A careful choice of the plasma viewing position and appropriate dilution factors led to accurate and sensitive determinations of REEs. The proposed procedure using external calibration presented good accuracy and sensitivity. 650 $arare earth elements 653 $aGeological material 653 $aSpectrometry 700 1 $aMACHADO, R. C. 700 1 $aBARROS, J. A. V. de A. 700 1 $aVIRGILIO, A. 700 1 $aSCHIAVO, D. 700 1 $aNOGUEIRA, A. R. de A. 700 1 $aNOBREGA, J. de A.